Q: What is the maximum sterilization temperature for the Bohrsägensystem?
The maximum sterilization temperature is ≤ 135 °C (improved models can withstand ≤ 155 °C). Exceeding 135 °C may lead to permanent motor damage.
Q: Can the device battery be sterilized under high temperatures?
No, the battery is not resistant to high temperatures and cannot be sterilized. Please ensure it is removed before any sterilization process.
Q: Why must the battery be removed from the handpiece after surgery?
Leaving the battery inside the tool leads to a slow automatic discharge, which can significantly shorten the overall lifespan of the battery.
Q: How should the cranial bur be maintained and cleaned?
It must be washed immediately after the surgery with a purger to prevent blood from sticking inside the bur, and then dried thoroughly. Additionally, ensure no water remains in the battery compartment of the handpiece after sterilization.
Q: How many times can a single cranial bur blade be used?
The cranial bur blade is made of flexible medical stainless steel and is susceptible to wear-and-tear. To ensure optimal performance and safety, it is highly recommended to use each blade for only one surgical procedure.
Q: What are the specifications for different drill attachments?
A normal bone drill holds drill bits under 6 mm. A reaming drill operates with high torque at approximately 300 rpm, and a hollow drill is used with K-wires or nails at approximately 600 rpm.